In the food industry, it is crucial to validate analytical methods for the testing of nutrients and components in processed food products, especially with the recent inclusion of warning labels. This paper introduces a novel methodology for determining sodium levels in fried corn-based snacks. The main objective of this study was to determine the reliability of the method in consistently delivering accurate results. The methodology involved a thorough validation of the flame photometry analytical method to precisely measure the sodium content in corn-based snack foods. The equipment results were linearized for the specific analyte, and its accuracy was determined by calibration curves at four control points using two sodium standards. One standard was standardized control, and the second was prepared by directly diluting the food matrix of interest. The limits of the equipment for sodium detection and quantification were determined. Accuracy was evaluated using experimental recovery percentages. The results of the method’s standardization were evaluated for 10 months by analyzing 150 corn-based snack samples and obtaining measurement data from stable sodium tests compared to the data reported on the packaging. The results indicated the accuracy and linearity of the method through a linear regression curve and analysis of variance, as well as homogeneity and homoscedasticity by applying the t-statistic and the Cochran Q test. Finally, the conclusion is that the method provides fast, safe, reliable, and accurate results. The sample analyzed is in compliance with the regulations for low-sodium foods without a warning label.
Keywords: Accuracy determination, calibration curves, corn-based snacks, flame photometry, sodium content, system validation.
En la industria alimentaria es crucial validar los métodos analíticos para el análisis de nutrientes y componentes en productos alimenticios procesados, especialmente, con la reciente inclusión del sistema de etiquetado frontal de alimentos. En este artículo se presenta una metodología novedosa para determinar los niveles de sodio en pasabocas fritos a base de maíz. Esta investigación tuvo como objetivo determinar la fiabilidad del método a la hora de proporcionar resultados precisos de forma consistente. La metodología consistió en una validación exhaustiva del método analítico de fotometría de llama para medir con precisión el contenido de sodio en pasabocas a base de maíz. Los resultados del equipo se linealizaron para el analito específico, y su precisión se determinó mediante curvas de calibración en cuatro puntos de control utilizando dos estándares de sodio. Un estándar era un control normalizado, y el segundo se preparó diluyendo directamente la matriz alimentaria de interés. Se determinaron los límites del equipo para la detección y cuantificación de sodio. La precisión se evaluó utilizando porcentajes experimentales de recuperación. Los resultados de la normalización del método se evaluaron durante 10 meses analizando 150 muestras de aperitivos a base de maíz y obteniendo datos de medición de las pruebas de sodio estable comparados con los datos indicados en el envase. Los resultados indicaron la precisión y linealidad del método mediante una curva de regresión lineal y un análisis de varianza, así como la homogeneidad y homocedasticidad aplicando el estadístico t y la prueba Q de Cochran. Finalmente, se concluye que el método proporciona resultados rápidos, seguros, fiables y precisos. Adicionalmente, se encontró que la muestra analizada cumple la normativa sobre alimentos bajos en sodio sin etiqueta de advertencia.
Palabras clave: Determinación de precisión, curvas de calibración, aperitivos de maíz, fotometría de flama, contenido de sodio, validación del sistema.
Salt (NaCl) is a crucial component in processed foods. Salt sodium is the main cation in extracellular fluids in animals and humans, like blood plasma
The salt-containing food products industry needs to obtain information quickly and easily. There must be methods to determine the safety of the products, and these methods need to be efficient, safe, and highly reliable. Therefore, choosing an accurate method for determining sodium content has become important for food and beverage manufacturers
On the contrary, Mohr titration involves the quantification of halides, particularly chlorides, utilizing a standardized solution of silver nitrate (AgNO3) as the titrant and potassium chromate (K2CrO4) as the indicator
The flame photometry method is based on the characteristic light emission that occurs when a gaseous atom in the flame is excited by heat
The flame photometry equipment is composed of an atomizer that disperses the sample generating a cloud that when in contact with the flame generates a signal, this signal that leaves the atomizer contains the information of the analyte and other radiations produced that pass to a monochromator
Flame photometry, a classic analytical technique, used in non-alimentary contexts, is widely used in the evaluation of alkali and alkaline soil metals, such as sodium, potassium, calcium, and magnesium. In the context of semiconductor sensors
In the food industry, validating other analytical methods is crucial to ensuring safe, accurate, and reliable results for nutrient testing and components in processed food products. But it is also important how easy and fast these methods should be.
The flame photometer has been utilized in some food-related applications. For instance, it has been employed to analyze minerals and heavy metals in Syzygium cumini fruit from different regions in Bangladesh
In general, few previous studies have used this technique for food characterization. Specifically, in products such as extruded and fried snacks made from corn, where the sodium level can vary considerably due to the manufacturing processes and the reactions of ingredients used at high fried temperatures, it is important to rely on techniques such as flame photometry. Based on this, this study aimed to validate the flame photometry method for determining sodium content in extruded and fried corn-based snack matrices for the first time.
A Digimed® flame photometer (DM-64-4E, Sao Paulo, Brazil) was used for the present study; measurements were performed based on the recommendations of the manufacturer, precision level of 1 % or 0.1 ppm, response time 8 s, volume per analysis 2 mL. The photometer is programmed to aspirate 7.00 ± 0.20 mL min-1 of the standards and samples. An analytical balance (±0.0001 g, Sartorius Entris® II, Advanced BCE224I-1S, Goettingen, Germany), was used for weighing the samples. Micropipettes (Transferpette®, D-10000, Germany) with adjustable volumes were utilized during the standards preparation. A food processor (Oster®, FPSTFP1455, China) was used to homogenize the samples before testing.
The glassware was washed with deionized water before each analysis and all the glassware was washed weekly with 10 % (v/v) HNO3 for 30 min and then rinsed with plenty of deionized water to remove traces of acid.
High-purity deionized water obtained from the Elix Advantage® 3/5/10/15 F3DA70491A system (resistivity 0.275 µS cm-1) was used to prepare the analytical standards and samples. The standard reference solution for the calibration curves was a sodium solution of Digimed® DM-513KC-64-A (Sao Paulo, Brazil).
Samples were collected monthly at 15 samples per month for 10 months. The samples were 30 g of fried corn-based snack matrixes (SNK-MN).
The SNK-MN samples were deposited in the food processor and the product was homogenized at maximum power at 30 s intervals, ensuring that the entire sample was completely homogenized. A clean and dry beaker was tared in the analytical balance until 100 g of sample was obtained. From the initial sample, 30 g were selected, using clean and dry watch glass. The sample of 30 g was placed in a balloon and was made up to 250 mL with constant agitation. The sample was allowed to settle for 3 min and 10 mL was withdrawn with the pipette. The 10 mL of diluted sample was again volumed to 100 mL, and stirred constantly for 5 min, from the second dilution 10 mL were removed and transferred to a beaker to feed the aspiration system of the flame photometer. The photometer was previously turned on and calibrated for sodium reading, the sample was atomized and the reading in ppm Na was generated from 0.00 ppm to 100 ppm. This procedure was performed in triplicate (n = 3) and with an analytical blank. The official AOAC method 963.13-1964,
To determine the concentration in mg Na / 100 g in the finished product from the flame photometer reading data in ppm was used (1)
\[ mg Na = \left(\frac{X_1* \frac{V_1}{V_2}*V_3}{M}\right) * (100g) \tag{1} \]
Where X1 is sodium result, photometer measurement (mg/L), V1 is volume dilution (L), V2 is volume of the aliquot taken to make the dilution (L), V3 is aliquot volume (L), and M is sample mass of finished product (g).
The data obtained in the experimental tests is quantitative. The data was handled using descriptive statistics. The standard calibration curve was performed based on specific concentration standards of 2, 5, 10, 20, 30, 50, 80, and 100 ppm of standard solution, with wavelength-specific readings for sodium in 589 nm.
The linearity was evaluated by elaborating two calibration curves using the standard standards and the solution of the SNK-MN samples at dilutions of 0.02, 0.05, 0.10, 0.20, 0.30, 0.50, 0.80, 1.00 % (v/v) and determining the linearity of the photometer reading by implementing linear regression statistics, analysis of variance for the regression curve and statistical t-test. The limits of quantification (LOQ) and detection (LOD) were calculated using the equation of the linear regression line
For the precision of the method, four controls corresponding to calibration points for sodium were taken from the standard calibration curve with concentrations of 10, 30, 50, and 80 ppm of the standard solution, performing four measurements for each standard. A “Q test” of Cochran was conducted to assess homogeneity among four groups of calibration curve measurements
\[ Q = \frac{(k-1)*\sum_{i=1}^{k} = 1*(X_i - \bar{X})^2}{\sum_{j=1}^{n}*(X_{ij} - X_i)^2} \tag{2} \]
Where Q is Cochran's statistic, k is the number of groups, n is the total number of observations, Xi is the mean of the variances of groups, is the overall mean of the variances, and Xij is the variance of the jth element of the ith group. Equation (2) was simplified and converted into (3)
\[ Q = \frac{(K-1)\text{SSbetween}}{\text{SSwithin}} \tag{3} \]
Where k is the number of groups, \(*\sum_{i=1}^{k}*(X_i - \bar{X})^2\) is expressed as the sum of squares between the groups (SSbetween), \(\sum_{j=1}^{n}*(X_{ij} - X_i)^2 \) is expressed as the sum of squares within the groups (SSwithin).
For the data stability test, the X-S control plot was used to evaluate the variability of the process versus its mean (X) and standard deviation (S), Determining the upper and lower limits, as well as the central limits, using 15 observations for 10 months with a total of 150 data. To determine X, (4), (5) and (6) were used:
\[ LCS = \bar{\bar{X}} + 3 \frac{\bar{S}}{c4\sqrt{n}} \tag{4} \]
\[ \text{Central limit} = \bar{\bar{X}} \tag{5} \]
\[ LCI = \bar{\bar{X}} - 3 \frac{\bar{S}}{c4\sqrt{n}} \tag{6} \]
To determine S, (7), (8) and (9) were used:
\[ LCS = \bar{S} + 3 \frac{\bar{S}}{c4} \sqrt{1-c4^2} \tag{7} \]
\[ Central limit = \bar{S} \tag{8}\]
\[ LCI = \bar{S} - 3 \frac{\bar{S}}{c4} \sqrt{1-c4^2} \tag{9} \]
Where \( \bar{\bar{X}} \) is the average of the average of the observations, \(\bar{\bar{S}}\) is the average of the standard deviations, C4 is a specific variable for the creation of control charts, and n is the number of observations.
For the accuracy determination, the recovery percentages were calculated and compared to the theoretical value of the standard solution at the calibration points in sodium (ppm).
Table 1 shows the calibration data and recovery percentages. For the calibration curves, an average percentage recovery R of 88 % was obtained, which is within the range of 80 % and 120 % as determined by
| Theoretical concentration (ppm) | Experimental concentration (ppm)* | Recovery (%) |
| 0.02 | 0.022 ± 0.0025 | 108 % |
| 0.05 | 0.053 ± 0.0036 | 106 % |
| 0.10 | 0.113 ± 0.0144 | 113 % |
| 0.20 | 0.212 ± 0.0335 | 106 % |
| 0.30 | 0.281 ± 0.0915 | 94 % |
| 0.50 | 0.493 ± 0.0815 | 99 % |
| 0.80 | 0.792 ± 0.0695 | 99 % |
| 1.00 | 0.981 ± 0.0583 | 98 % |
A calibration graph with practically overlapping data was obtained as shown in Figure 1 with a coefficient of determination of 0.993 for standard calibration and 0.9985 for saline calibration (Data in Table 1s of the supplementary material). These coefficients demonstrate the fit of the experimental data versus the photometer reading points for the concentrations analyzed and the absence of systematic reading error
Curve regression statistics indicate a positive correlation in the linear regression model with a 0.999 correlation coefficient. Based on the above, the closeness of the data obtained by flame photometry versus the measurement points established for the analyzed concentrations of sodium in ppm can be demonstrated. The coefficient of determination indicates the fit of the regression line against the analyzed data obtaining a value of 0.998; the determined standard error obtained is 0.0142 demonstrating the fit of the model
The analysis of variance obtained an F value of 4545.87 and a critical F value of 4.15x10-11 indicating that the regression model has a high statistical significance since the critical F value is lower than the calculated F, evidencing the importance of the relationship between variances for the reading data in the flame photometer for sodium in ppm and the correlation of the values, a strong relationship is determined that is statistically significant between the theoretical concentration in ppm and the experimental measurements, determining the linearity of the method as demonstrated by the high F value and the low P value for the theoretical concentration of measurement; the upper and lower limits of the intercept include 0 and therefore the line is compatible to pass through the intercept (Table 2S of the supplementary material).
The agreement between the experimental measurements and the theoretical calibration points in the flame photometer is evidenced by observing an overlap in the points of the fitted regression curve (Figure 1), demonstrating precision in the method and reliability of the obtained results in the determination of sodium by flame photometry of the intensity measured versus the concentration.
The experimental values often do not match the mathematical model, leading to differences between observed and predicted values

When performing the statistical t-test (Table 3S of the supplementary material) it was determined that there is no statistical difference between the concentrations of the two samples; the mean and variance of the two samples are comparable and the statistical analysis does not indicate sufficient evidence to reject the null hypothesis of hypothetical difference of the means different from zero, this shows that the flame photometry method is efficient and also demonstrates the accuracy it has for the determination of sodium in ppm in the analyzed solutions.
The test statistical analysis of Cochran was used to validate the homogeneity of variances demonstrating homoscedastic behavior using a probability of 0.05. A Q value of 0.0000221 was obtained (Table 2), which is below the critical value analyzed by the Chi-square table (in the original Cochran´s paper) with (k-1) which is 7.815 for three degrees of freedom
| Calibration curve | Curve data | Media | Total mean | SSbetween | SSwithin | Q | |||
| 1 | 0.010 | 0.299 | 0.493 | 0.802 | 0.401 | 0.401 | 0.00003859 | 1.7434875 | 0.0000221 |
| 2 | 0.010 | 0.301 | 0.502 | 0.796 | 0.402 | ||||
| 3 | 0.010 | 0.298 | 0.491 | 0.794 | 0.398 | ||||
| 4 | 0.010 | 0.296 | 0.494 | 0.809 | 0.402 | ||||
To determine the stability of the samples during the study, a control X plot was made to determine the behavior versus the mean of the data (Figure 3), showing a distribution of the data above the mean control line and a peak outside the established limits, indicating that the experimental control data are stable for the mean during the 10 months of testing for the 150 samples analyzed.

Figure 4 for the control of the stability of the samples shows that the process is controlled for the standard deviation demonstrating the stability of the 150 samples with scattered data on the control line demonstrating a controlled trend of the data without exceeding the maximum and minimum limits established. Figure 4 and Figure 5 demonstrate the stable behavior of the product analyzed during the ten months of study and follow-up, evidencing control over the process and stability of the product, generating reliability of the data obtained in the sodium analysis. The concentration in mg Na/100 g of the finished product reported on the packaging is 269. For the data analyzed to the reported value of 269 mg Na/100 g, an average sample mean of 269.53 was obtained using (1) and an average standard deviation of 1.87, determining, as shown in the X-S graphs, stability in the samples over time and experimental sodium reading values with low deviation in comparison with the value reported in the product. The results are lower than the limits established by Resolution 2013 of 2020 in Colombia, which establishes the technical regulation that defines the maximum sodium content of processed foods.

The results obtained show that the analytical method of determination by flame photometry offers safe, precise, reliable, and accurate data for the determination of sodium in corn-based food matrixes. Preparing sample dilutions reduces analysis time.
Using the implemented statistics, it was possible to determine that the parameters of linearity, accuracy, specificity, and repeatability are acceptable for quantifying sodium in the snack sample by flame photometry with a limit of detection LOD of 0.015 ppm and a limit of quantification LOQ of 0. 0526 ppm for sodium. Similarly, compliance with the acceptance criteria is evident in all the tests performed. This guarantees the efficiency and effectiveness of implementing the flame photometry method for determining sodium in corn-based snack foods. However, this test is not limited to this type of food matrix, as it could also be applied for alkali metals detection in dairy products, meats, beverages, and other processed foods.
The accuracy of the method was assessed by comparing the recovery percentages with the theoretical values of the standards utilized. The findings revealed a strong correlation between the theoretical and experimental concentrations, indicating the precision and accuracy of the flame photometry method in determining sodium. Determining sodium in processed foods is essential for compliance with high sodium regulations.
The analyzed sample complies with regulations for low-sodium foods without a warning label. However, the main limitation of the method is the requirement for frequent calibration since, to maintain accuracy, it is necessary to calibrate the equipment constantly with known sodium standards and with the specific standards for the equipment, which have an average cost of USD 3700 and a useful life of six months.
This research received no specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
The authors confirm that they have no conflict of interest for this manuscript.
Jhojan Mauricio Madrid-Molina: Conceived the Idea, Collected Data, Contributed Data Processing or Analysis Tools, and Wrote the Article